Section 8-23 - Replacement of the COM Express
2) Type “maru” for password.
3) After successfully entering BIOS, verify the following parameters
Table 8-24 Configuring BIOS.
4) Change date/time parameters as needed, and save the changes.
5) Reboot the console.
8-23-6
Note on EEPROM Replacement
NOTE:
EEPROM on MCB contains VPD (Vital Product Data) for SOM (ComExpress). This VPD contains
information such as system serial number to be matched to Software License Key, or Board revision
number that is cross referenced to system software for compatibility.
Service personnel may require to initialize EEPROM or to modify EEPROM contents for proper
operation of the system.
Figure 8-156 EEPROM on MCB Board
TAB
top menu
Main
System Time
System Date
Left bottom key of trackball to
enter BIOS and select menu and set
or change the value in BIOS
Right bottom key of trackball
for ‘ESC’ in BIOS
EEPROM on MCB Board Memory