Chapter 4 - Functional Checks
4-24
4-3-6-6
Gel Warmer Test
1) Insert Gel Bottle into Gel Warmer.
2) Power ON the warmer.
3) Verify Gel gets warm.
NOTE:
Depending on environment, Gel requres 30 minute or more to warm up.
4-3-6-7
Single CWD
1) Connect P8D probe. Select vascular application like “Carotid”. And start scan.
2) Verify that CW timeline image appears.
4-3-7
Mode Transition Checks
4-3-7-1
General Information
1) [R1.X.X] Select UTILITY -> SYSTEM -> User Configurable Key.
[R2.X.X or later] Select UTILITY -> APPLICATION -> Settings.
2) [R1.X.X] Select "M" in 2 list of "Mode Rotary Line Key".
[R2.X.X or later] Select "M" in 2 list of "User Configurable Mode Rotary".
3) [R1.X.X] Select "PW" in 3 list of "Mode Rotary Line Key".
[R2.X.X or later] Select "PW" in 3 list of "User Configurable Mode Rotary".
4) [R1.X.X] Select "CF" in 5 list of "Mode Rotary Line Key".
[R2.X.X or later] Select "CF" in 5 list of "User Configurable Mode Rotary".
5) Press “Save”.
6) Select UTILITY -> SYSTEM -> User Configurable Key.
7) Select “B-Flow” in 4 list of “User Defined Hard Key”.
8) Select "Reset" in 3 list of "User Defined Touch Button".
Содержание LOGIQ P9
Страница 2: ......
Страница 11: ...GE LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 SERVICE MANUAL ix ZH CN KO ...
Страница 40: ...GE LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 SERVICE MANUAL xxxviii Table of Contents ...
Страница 184: ...GEDRAFT LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 DRAFT JANUARY 24 2019 SERVICE MANUAL Chapter 4 Functional Checks 4 44 ...
Страница 601: ...This page was intentionally left blank ...