Chapter 7 - Diagnostics/Troubleshooting
7-9
7-4-3
System Warning/Errors and Logs
7-4-3-1
Temperature warning
7-4-3-1-1
Cause
CPU Temperature exceeded threshold (104°C).
7-4-3-1-2
What to do
1) Shut down the system and Clean the filter.
7-4-3-2
Temperature exceeded threshold
7-4-3-2-1
Cause
CPU Temperature exceeded threshold (100°C) and therefore, system must be shut down immediately.
7-4-3-2-2
What to do
1) Shut down the system and Clean the filter.
7-4-3-2-3
Log message of the high temperature
(Example)
Error ; acqHouseKeeping(****); CPU Temperature Sensor BEP_CPU_TEMPERATURE = 105.00
degrees C is outside tolerance limits. UpperLimit = 104.00 LowerLimit = -10.00