Section 8-17 - Replacement around Nest Box
2) Remove MCWD board from MCB assy. Refer to following figure.
NOTE:
Be careful not to damage CWD connector.
Figure 8-126 Removal of MCWD board.
3) Remove Stud on MCB assy and carefully lift up the MCB assy. Refer to following figure.
Figure 8-127 Stud position on MCB board.
MCWD board assy