Section 5-11 - Common Service Desktop (R3 only)
To view the Class M license information
:
1.) Insert the SSA key.
2.) Navigate to
Utility (second page) >
Service > Utilities > SSA License
.
3.) Under
Class M key Details
, view the values. For example, the
SSO ID
for the user assigned to the
SSA key.
If
Not Available
displays for all of these values, the SSA key is not validating.
To restore an SSA key that is not validating
:
1.) Remove the SSA key from the LOGIQ P9/P7.
2.) If open, close the Service desktop.
3.) Restart the LOGIQ P9/P7.
4.) Once the LOGIQ P9/P7 has restarted, plug in the SSA key.
5.) Navigate to
Utility (second page) >
Service > Utilities > SSA License
.
6.) Click
Restore SSA
.
7.) Check to see if the SSA key validates.
For more information, see
:
•
•
SSA key on page 7-94