Chapter 4 - Functional Checks
4-28
4-3-7-4
Pulse Doppler Mode
Figure 4-26
1) From Utility -> Imaging -> General, verify check on “Simultaneous”.
2) Press B Mode button.
3) Press PW Mode button. (Active PW Mode button on "User Configurable Mode Rotary".)
4) Verify that system scan B/PW duplex with the activated probe.
5) Revert to original setting in Utility -> Imaging -> General “Simultaneous”.
Figure 4-27 Pulse Doppler Mode
4-3-7-5
B/CF/PW Mode
1) Press B Mode button.
2) Press PW Mode Button.
3) Press CF Mode Button.