Section 5-11 - Common Service Desktop (R3 only)
To reload software
:
1.) If the drives are encrypted, obtain the recovery key.
2.) Navigate to
Utility (second page) > Service > Utilities > Disruptive Mode Utility
and set
Disruptive mode to On.
3.) Navigate to
Utility (second page) > Service > Utilities > Software Reload
.
4.) View the current version numbers.
5.) Click
Software Reload
.
6.) If you have a recovery key, select
I have a recovery key
and then click
Continue
.
7.) If you have a recovery key, select
I Agree
and then click
Continue Reload
.
For more information, see
:
•
CAUTION
RELOADING SOFTWARE WITHOUT A RECOVERY KEY WILL RESULT IN THE LOSS OF
PATIENT DATA.