10-10
Section 10-5 - System Maintenance
10-5-5
Physical Inspection
10-5-6
Optional
Diagnostic Checks
Optionally you can access the diagnostic software as described in Chapters 5 or 7.
View the error logs and run desired diagnostics.
10-5-6-1
View the logs
1) Review the system error log for any problems.
2) Check the temperature log to see if there are any trends that could cause problems in the future.
3) Check the Configuration Log; update if needed.
Table 10-9 Physical Checks
Step
Item
Description
1
Labeling
Verify that all system labeling is present and in readable condition.
2
Scratches & Dents
Inspect the console for dents, scratches or cracks.
3
LCD Monitor Display
Inspect the LCD Monitor Display for scratches and raster burns.
Verify proper operation of Contrast and Brightness controls.
4
Control Panel and
Keyboard
Inspect the Control Panel and Keyboard. Note any damaged or missing items.
(Replace faulty components, as required).
Verify proper operation of Control Panel backlighting and TGC sliders.
5
DVD Drive
Clean the drive head and media with the vendor-supplied cleaning kit.Advise the user to repeat
this often, to prevent future problems. DVDs/CDs must be stored away from dust and cigarette
smoke. Do not use alcohol or benzene to clean the drive
6
Wheels & Brakes
Check all wheels and casters for wear and verify operation of foot brake, to stop the unit from
moving, and release mechanism. Check all wheel locks and swivel locks for proper operation.
7
Cables & Connectors
Check all internal cable harnesses and connectors for wear and secure connector seating.
Pay special attention to footswitch assembly and probe strain or bend reliefs.
8
Power Cord
Check the power cord for cuts, loose hardware, tire marks, exposed insulation or other
deterioration, and verify continuity. Tighten the clamps that secure the power cord to the unit and
the outlet plug to the cord.
9
Shielding & Covers
Check to ensure that all EMI shielding, internal covers, air flow panels and screws are in place.
Missing covers and hardware could cause EMI/RFI problems during scanning.
10
Peripherals
Check and clean the peripherals according to the manufacturer’s directions.
To prevent EMI or system overheating, dress the peripheral cables inside the peripheral cover.
11
External I/O
Check all connectors for damage and verify that the labeling is good.
12
Op Panel Lights
Check proper operation of all control panel key illuminations (flash once during system start-up).
Содержание LOGIQ P9
Страница 2: ......
Страница 11: ...GE LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 SERVICE MANUAL ix ZH CN KO ...
Страница 40: ...GE LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 SERVICE MANUAL xxxviii Table of Contents ...
Страница 184: ...GEDRAFT LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 DRAFT JANUARY 24 2019 SERVICE MANUAL Chapter 4 Functional Checks 4 44 ...
Страница 601: ...This page was intentionally left blank ...