Section 10-8 - Electrical Safety Tests
10-8-5
Enclosure Leakage Current Test
10-8-5-1
Definition
This test measures the current that would flow in a grounded person who touched accessible non-
protectively-earthed parts of equipment if the ground wire were to break. The test verifies the isolation
of the power line from the chassis.
The meter is connected from accessible non-protectively-earthed parts of the case to ground.
Measurements should be made with the ground open and closed, with power line polarity normal and
reversed, and with the neutral open and closed. Record the highest reading.
10-8-5-2
Generic Procedure
The test verifies the isolation of the power line from the enclosure. The testing meter is connected from
accessible non-protectively-earthed parts of the case to ground. Measurements should be made with
the ground open and closed, with power line polarity normal and reversed, and with the neutral open
and closed. Record the highest reading of current.
NOTE:
Color of Power outlet cable
i) USA
and
Canada
Hot: Black
Neutral: White
Ground: Green or Green-yellow
ii) Others
Hot: Brown
Neutral: Blue
Ground: Green-yellow
DANGER
!! DANGER:
Electric Shock Hazard.
When the meter’s ground switch is OPEN, DO NOT touch the unit!
CAUTION
!! CAUTION:
Equipment damage possibility. Never switch the Polarity and the status of Neutral when the unit
is powered ON. Be sure to turn the unit power OFF before switching them using the POLARITY
switch and/or the NEUTRAL switch. Otherwise, the unit may be damaged.
Figure 10-18 Set Up for Test of Enclosure Leakage Current, UL60601-1/IEC 60601-1 Clause 19
Содержание LOGIQ P9
Страница 2: ......
Страница 11: ...GE LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 SERVICE MANUAL ix ZH CN KO ...
Страница 40: ...GE LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 SERVICE MANUAL xxxviii Table of Contents ...
Страница 184: ...GEDRAFT LOGIQ P9 P7 DIRECTION 5604324 REVISION 11 DRAFT JANUARY 24 2019 SERVICE MANUAL Chapter 4 Functional Checks 4 44 ...
Страница 601: ...This page was intentionally left blank ...