3D Systems, Inc.
97
p/n: 15-D100, rev. D
Inlet
valve
Outlet
valve
5. Unscrew the philips screw on the oxygen sensor and
disconnect the sensor from the outlet valve.
6. Remove the KF-clamps on inlet and outlet valves of the process filter (see images below).
7. Roll out the process filter cart to a safe and secure location.
8. Disconnect and remove the KF50 clamp between the
outlet valve and the process filter (see image to the right).
Remove the filter outlet valve assembly and the centering
ring with stand-off. The centering ring may stick slightly
to the filter housing. If needed, pry it loose with a plastic
tool (such as a scraper).
9. As you lift the centering ring away from the
filter,simultaneously check to make certain the spring-
loaded valve has closed.
WARNING: THE FILTER OUTLET PORT MUST BE
CAPPED IMMEDIATELY AFTER THE FILTER OUTLET
VALVE ASSEMBLY IS REMOVED.