15-3
15
IM 701830-01E
Specifications
Vertical/horizontal axis setting functions
Item
Specifications
Channel ON/OFF
Independent ON/OFF for each channel (CH1 to CH16 and 32-bits extended logic input).
Vertical-axis Expand/reduce
Axis can be expanded or reduced.
Input filter
Can be independently set for each module
Display inversion
Invert the waveform display with respect to the vertical position.
Linear scaling
Set scaling coefficient, offset, and unit separately for each channel (CH1 to CH16).
Roll mode
100 ms/div to 200 ks/div, and the sample rate is 100 kS/s or less (excluding a portion of the waveform
acquisition methods.)
Analysis Functions
Item
Specifications
Cursor measurement function
Allows selection of cursor type from Marker, Horiz, Vert, H&V, and UserDefine.
Automatic measurement of
Measures the following waveform parameters automatically.
waveform parameters function
P-P, Max, Min, High, Low, Ave, Rms, Amp, StdDev, +Oshot, -Oshot, Rise, Fall, Freq, Period,
+Duty, -Duty, +Width, -Width, RDelay, FDelay, Int1TY, Int2TY, Int1XY, Int2XY and Histo
Computing functions*
Addition, subtraction and multiplication, binary codes, phase shift, power spectrum, DA
Use defined computation
The following operators can be combined in any way to set the computing equation.
(Optional)
ABS, SQR, LOG, EXP, BIN, PWHH, PWHL, PWLH, PWLL, PWXX, ATAN, MEAN, DIF, DDIF,
INTG, IINTG, PH, HLBT, P2, P3, FILT1, FILT2, LS-, RS-, PS-, PSD-, TF-, CH-, PHASE, REAL,
IMAG, LOGMAG, MAG, F1, F2
* The maximum record lengths are 400 kW (using Math1 and Math2) or 100 kW (using Math1 to Math8).
Screen Data Output Functions
Item
Specifications
Built-in printer/Printer with
Output the hard copy of the screen in Normal or Long format.
Centronics interface
GP-IB/RS-232 interface
Output data formats : HP-GL, PostScript, ThinkJet, TIFF, BMP
Floppy disk, Internal HDD,
Output data formats : HP-GL, PostScript, ThinkJet, TIFF, BMP
SCSI device
Data Storage Functions
Item
Specifications
History memory
Up to 1000 acquisition data records can be automatically stored (for /M1, /M2 or /M3 model).
Floppy/Internal HDD/
Save and restore waveform data, settings, other data.
SCSI device
Other Functions
Item
Specifications
Initialization function
Automatically resets key settings to the factory settings.
Auto set-up function
Automatically sets key settings to the optimum values for the input signals.
Calibration
The calibration items : Ground level offset, A/D converter gain
Environment setting function
Allows setting of screen color, date/time, message language, click sound ON/OFF.
COMP signal output function
Outputs a square calibration waveform signal (1 Vp-p
±
10%, 1 kHz
±
10%) from the COMP terminal
on the front panel.
Overview function
Shows system configuration.
Self test function
Allows memory test, key test, FDD, Internal HDD, and printer test.
Help function
Displays help about settings.
15.5 Functions