6-2
IM 701830-01E
6.2
Selecting a Channel for Setting Trigger
Hysteresis and Trigger Level
Function
Selects the channel (trigger source) for setting parameters such as trigger hysteresis and trigger
level.
If you use a simple trigger, only the selected channel serves as the trigger source. If you use an
enhanced trigger, the trigger source is generated using all channels selected within the “Set
Pattern” function.
The following traces are available for trigger source.
• CH1 to CH8 (CH2, CH4, CH6)
• LOGIC1 and LOGIC2 (32-bit extended logic input)
• EXT (the TRIG IN terminal on the left side)
• LINE (commercial power supply signal)
Note
The items to set are different when CH1 to CH16 is selected and when LOGIC1, LOGIC2, EXT, or LINE
is selected.
• CH1 to CH16 : hysteresis (Section 6.3), trigger level (Section 6.4), slope (simple
(for voltage/temperature module)
trigger only, Section 6.8)
• CH1 to CH16
: slope (simple trigger only, Section 6.8), Logic Bit No. (simple
(for logic input module)
trigger only, Section 6.8), Set Pattern (enhanced trigger only,
Section 6.9 to 6.14)
• LOGIC1, LOGIC2
: slope (simple trigger only, Section 6.8), Logic Bit No. (simple
trigger only, Section 6.8)
• EXT
: trigger level (Section 6.4), slope (simple trigger only, Section 6.8)
• LINE
: There are no trigger hysterisis/trigger level settings
Operating Procedure
1.
Press the
SIMPLE
key or the
ENHANCED
key.
2.
Press the “
Source
” soft key to display the trigger source menu.
3.
Select the desired trigger source from channels “
CH1
” to “
CH16
,” “
LOGIC1
,
*1, *2
”
“
LOGIC2
,
*1, *2
” “
EXT
*” and “
LINE
*” using the jog shuttle.
Channels which do not have input modules installed are not displayed on the menu.
*1 You can select this channel only when the optional 32-bit extended logic input (/N1) is
installed.
*2 Available only on simple trigger.
Simple trigger mode
Enhanced trigger mode