Functions
B>Time, B<Time, B Time out (Enhanced Trigger)
→
Page 6-17.
A trigger is activated on the falling or rising edge of the pulse when the pulse width exceeds (or
drops below) the preset time. In the case of a “Time out” trigger, a trigger is activated when the
preset time elapses. If multiple signals are used, a trigger is activated by the AND of the
multiple signals.
When B > Time, CH1 = H, Time = 4
µ
s
Trigger
5
µ
s
1
µ
s
CH1
When B < Time, CH1 = H, Time = 4
µ
s
Trigger
5
µ
s
1
µ
s
CH1
When B Time Out, CH1 = H, Time = 4
µ
s
Trigger
5
µ
s
1
µ
s
CH1
4
µ
s
Window Trigger (Enhanced Trigger)
→
page 6-19.
A certain voltage range (window) is set and a trigger is activated when the trigger source level
enters this voltage range (
) or exits from this voltage range ( ).
Width
trigger is activated
trigger is activated
Window
Window
trigger is activated
Center
Width
Center
Timer Trigger (Simple Trigger)
→
page 6-21.
The trigger is activated at the specified interval from the specified time.
Trigger
Trigger
Trigger
Trigger
Specified time
Specified
Interval
Specified
Interval
Specified
Interval
1.3 Setting a Trigger