5.11 Inverting a Waveform
(For High-Speed Isolation/High-Speed/High-Resolution, High-Voltage, Isolation/High-
Resolution, Isolation Modules)
≡
For a description of this function, refer to page 1-5.
≡
Function
Relevant channels
The input signals to channels CH1 to CH16 can be inverted independently of one another.
Inversion is performed about the vertical position.
Points for attention
• If you change the invert display setting while the waveform acquisition is stopped, the
display does not change. The waveform will be displayed inverted the next time you start the
acquisition.
• Cursor measurements, automatic measurement of waveform parameters, and computation
functions do not give correct results if the waveform is inverted and the offset is non-zero.
• You can not invert the waveform while realtime printing or realtime recording. The
waveform is inverted when you stop the waveform acquisition.
Operating Procedure
1.
Press one of the keys
CH1
to
CH16
to select the desired channel.
2.
Press the “
Next 1/2
” soft key.
3.
Press the “
Invert
” soft key until “
ON
” is selected.
→