14-2
IM 701830-01E
14.2 Error Messages and Corrective Actions
Error Messages
An error message may appear during operation. This section describes the meaning of each
message and the corrective action to be taken. Messages can be displayed in either Japanese or
English (refer to page 13-3). If the message indicates that maintenance service is required,
contact your nearest YOKOGAWA representative listed on the back cover of this manual.
In addition to the messages given below, there are also some other communications related error
messages (0 to 500). These messages are described in the Communication Interface User’s
Manual.
Error in Execution
Code
Error Description
Corrective Action
Reference Page
18
The instrument is set to key protect mode.
Press the PROTECT key
13-7
Press the PROTECT key to exit from this mode.
601
Invalid file name or SCSI ID.
Check filename or SCSI ID.
11-18
602, 603
No SCSI device or no media inserted.
Check the SCSI device connection and the
11-3, 11-8
SCSI ID, and make sure that the storage
medium is inserted (if applicable).
604
Media failure.
Check the storage medium.
–
605
File not found.
Check the filename and the storage medium.
Chapter 11
606
Media is protected.
Set the disk’s (medium’s) write protect switch
–
to OFF.
607
Media failure.
Check the storage medium.
–
608 to 610 Invalid file name or SCSI ID.
Check filename or SCSI ID.
11-18
611, 612
Media full.
Delete unnecessary file(s) or use another disk.
11-8
For the necessary capacity (bytes), refer to the
11-11
reference page numbers.
613
File system failure.
Check using another disk.
–
If the same message still appears, maintenance
service is required.
614
File is protected.
Delete the file’s “*” (write disable mark).
11-15
615
Physical format error.
Reformat the medium. If the same error occurs,
11-4
the instrument is probably unable to execute a
format on this medium.
616 to 620 File system failure.
Check using another disk.
–
If the same message still appears, maintenance
service is required.
621
File is damaged.
Check the file.
–
622 to 641, File system failure.
Check using another disk.
–
656 to 663
If the same message still appears, maintenance
service is required.
642
No media exists in SCSI device.
Check that the storage medium is correctly
11-7
inserted in the SCSI device.
646 to 653 Media failure.
Check the medium.
–
654
Media failure.
Use a floppy disk of a format supported by the
11-4
instrument.
665
Cannot load this file format.
Files stored on other models (DL/AG series)
–
cannot be loaded.
666
File is now being accessed. Wait a moment.
Execute after access is made.
–
667
Cannot be executed while running.
Press the
START/STOP
key to stop acquisition.
4-6
668
Cannot find ‘.HDR’ file.
Check the file.
11-4
669
Cannot find ‘.INF’ file.
Check the file.
11-4
670
Displayed channel does not exist.
Turn ON the display of the appropriate channel.
5-1
671
Save data not found.
Check for presence of data.
11-8
672
This model does not have the SCSI interface.
–
2
673
SCSI controller failure.
Maintenance service is required.
–
680
Illegal printer head position.
Set the release arm to the “HOLD” position.
10-2
681
Paper empty.
Load a roll chart.
10-2
682
Printer over heat.
Turn OFF the power immediately.
–
Maintenance service is required.
683
Printer over heat.
Turn OFF the power immediately.
–
Maintenance service is required.
685
Printer time out.
Maintenance service is required.
–
686
Centronics printer error.
Turn the power of the printer from OFF to ON.
–
687
Centronics printer is off line.
–
–
688
Centronics printer is out of paper.
–
–