Operating Procedure
Selecting the computation mode
1.
Press the
MATH
key.
2.
Press the “
Mode
” soft key to display the math menu.
3.
Keep pressing the “
Mod
e” soft key until “
Binary
” is selected.
Setting the simultaneous binary computation type
4.
Press the “
Math1
” soft key to select “
State
”.
Setting the threshold level for each channel (Math1 only)
5.
Press the “
Next 1/2
” soft key.
6.
Press the “
Threshold Trace
” soft key to display the threshold level channel menu.
7.
Select one of the channels from “
CH1
” to “
CH16
” using the jog shuttle.
Channels which do not have input modules installed are not displayed on the menu.
8.
Press the “
Threshold Up/Threshold Lo
” soft key.
The upper threshold level can now be set using the jog shuttle.
9.
Turn the jog shuttle to set the upper threshold level (High).
10.
Repeat the above steps to set the lower threshold level (Low).
11.
Repeat steps 6 to 10 until the threshold level has been set for all desired channels.
→
Setting the D/A conversion computation
4.
Keep pressing the “
Math2
” soft key until “
DA
” is selected.
Selecting the D/A conversion method
5.
Press the “
Next
” soft key to display the “
Next 2/2
” menu.
6.
Press the “
DA Mode
” soft key to select either “
Normal
” or “
Comple
”.
→
9.7 Simultaneous Binary Computation and D/A Conversion of All Channels