4-3
4
IM 701830-01E
Common Operations
4.2
Initializing Settings
Function
The initialization function allows you to reset parameter values which have been set using panel
keys to the default (factory settings). This is very convenient when you have to cancel the
previous settings or when you have to restart measurement from the beginning.
Initialization
Initialization means resetting parameters to their factory setting values. For details on factory
settings, refer to the Appendix 5.
Settings which cannot be initialized
Date and time
GP-IB/RS-232 interface related settings
SCSI ID number
To cancel initialization
If you have performed initializatioE†by mistake, press the “
Undo
” soft key. This will restore
the previous settings used before the initialization was performed.
The “
Undo
” operation remains available only while power stays on.
Initialization at power ON
Turning the power switch ON while holding down the
RESET
key will start initialization. This
also initializes settings relating to the GP-IB/RS-232 interface and SCSI ID number.
Operating Procedure
Performing initialization
1.
Press the
SETUP
key.
2.
Press the “
Initialize
” soft key to start initialization.
Canceling initialization
3.
Press the “
Undo
” soft key. This will restore the previous settings that were in effect before
initialization.
Note