14-3
14
IM 701830-01E
Troubleshooting, Maintenance and
Inspection
Code
Error Description
Corrective Action
Reference Page
689
Centronics printer is in use.
–
–
690
Cannot detect Centronics printer.
Turn ON the printer.
–
Check connectors.
701, 702
Cannot be executed while running.
Press the
START/STOP
key to stop acquisition.
4-6
703
There is no undo-data.
Undo is not possible since data which was
4-3
present before initialization and auto set-up is
4-4
now not available.
704
Cannot be executed while running.
Press the
START/STOP
key to stop acquisition.
4-6
707
Can’t start while data out. Wait a moment.
Wait until output is completed.
4-6
708
Can’t data out while running.
Press the
START/STOP
key to stop acquisition.
10-4, 10-7, 10-9
709
Can’t detect listener. Check GP-IB connector.
Check whether the GP-IB cable is connected
–
properly.
710
File not found.
Check the file.
–
711
Cannot access file while hard-copying.
Wait until output is completed.
–
712
Cannot compress this hardcopy image.
Set compression OFF.
10-12
713
Calibration failure.Failed CH: *
Maintenance service is required.
–
(The CH number which failed the calibration is
displayed at the * mark).
715
Realtime print possible for 500 ms/div or slower.
–
1-14, 7-9
716
Realtime store possible for 1 s/div or slower.
–
1-14, 7-11
717
Too many channels for the current T/div setting to
Decrease the number of channels by turning them
1-14, 7-11
realtime store.
OFF.
718
Too many channels for the current Store Length to Decrease the number of channels by turning them
1-14, 7-11
realtime store.
OFF or shorten the Store Length.
724
Balancing failed.
Check the strain gauge/strain gauge bridge
3-16, 5-19
Problematic channel: *
connections and the bridge voltage/gauge
* indicates the channel number of the channel
factor settings. Then, execute the balancing again.
for which automatic balancing failed.
725
Cannot be executed. Acquisition not active.
Press the
START/STOP
key to start acquisition.
4-6
Error in Setting (800 to 899)
Code
Error Description
Corrective Action
Reference Page
800
Illegal date-time.
Set the correct date and time.
3-6
801
Illegal file name.
The file name contains characters which are
4-2
not allowed or the file name is not a valid
MS-DOS file name. Enter another file name.
802
Max number of selection is 8 (except Histogram).
The number of waveform parameters which can
9-5, 9-8
be measured automatically is limited to a
maximum of eight.
803
Turn off accumulate-mode to enter history menu.
It is not possible to activate the history menu by
8-5
pressing the
HISTORY
key during accumulation.
804
Cannot change this parameter while running.
Press the
START/STOP
key to stop acquisition.
4-6
809
Cannot change edge for status other than X.
Set the state of the channel corresponding to
6-14
condition A to “X”.
813
Set Items in measure menu.
Set appropriate measurement items.
9-5, 9-6
814
Duplicated Name.
Change the label string.
8-9, 11-8
816
Cannot turn off with the current record length.
Reduce the record length.
7-1
Please Change record length.
817
Cannot change.
Go to X-Y menu and change the XTrace.
8-10
819
Cannot change when Channel Display is OFF
Set the channel display ON or make appropriate
5-1
or Math settings are invalid.
Math settings.
Sections 9.5 to 9.9
823
Cannot change while running.
Press the
START/STOP
key to stop acquisition.
4-6
824
Cannot change with the current acq mode
Set the acquisition mode to Normal.
7-4
Please change acq mode.
827
Illegal computing equation.
Input a correct computing equation.
9-19
830
Cannot set anything other than Low Pass for a
Change the Filter Type to another filter besides
9-20, 9-22
Gaussian filter.
Gaussian.
831
Cannot execute Auto Set-up on a channel which
–
4-4
has a logic input module installed.
832
Cannot set the edge on the logic module.
–
6-14
833
Cannot set logic module to B pattern.
–
6-10, 6-12, 6-17
834
Duplicate SCSI ID.
Set different ID numbers.
11-18
835
Cannot change settings while realtime printing or
Stop printing or recording.
7-10, 7-12
realtime recording.
836
Cannot change settings during Action On Trigger.
Stop the Action On Trigger.
13-5
837
Cannot set the channels which do not have modules –
–
installed.
838
This setting is invalid for channels where
Install the strain module into the corresponding CH. 3-8
a strain module is not installed.
14.2 Error Messages and Corrective Actions