5-24
IM 701830-01E
5.16 Using the Linear Scaling Function for the Strain
When
“
P1–P2
”
is selected
4.
Press the “
P1:X/P1:Y
” soft key and set the jog shuttle action to “
P1:X
.” “
P1:X
” is
displayed in the “
Get Measure
” soft key.
5.
Press the “
Get Measure
” soft key to read in the measured value into P1:X or set the value
with the jog shuttle.
You can change the setting a digit using the arrow keys.
6.
Press the “
P1:X/P1:Y
” soft key again and set the jog shuttle action to “
P1:Y
.”
7.
Set the scale value P1:Y with the jog shuttle.
You can change the setting a digit using the arrow keys.
8.
Press the “
P2:X/P2:Y
” soft key and set the jog shuttle action to “
P2:X
.” “
P2:X
” is
displayed in the “
Get Measure
" soft key.
9.
Press the “
Get Measure
” soft key to read in the measured value into P2:X or set the value
with the jog shuttle.
You can change the setting a digit using the arrow keys.
10.
Press the “
P2:X/P2:Y
” soft key again and set the jog shuttle action to “
P2:Y
.”
11.
Set the scale value P2:Y with the jog shuttle.
You can change the setting a digit using the arrow keys.
Set the unit
If necessary, press the “
Units
” soft key to display the keyboard, then input the name of the unit.
For the procedures on entering character strings, see section 4.1 “Entering Values and Character
Strings.”