15-13
15
IM 701830-01E
Specifications
15.22 Temperature Module (701860)
Item
Specifications
Number of input channels
1
Data update rate
Approx. 135 Hz
Input format
Isolated unbalanced
Sensor
Thermocouple
Measurement range/accuracy*
1
*
2
Type
Measurement range
Accuracy
K
-200 to 1300
°
C
±
(0.2% of r 1.5
°
C)
E
-200 to 800
°
C
except
±
(0.5% of reading
±
1.5
°
C)
J
-200 to 1100
°
C
T
-200 to 400
°
C
L
-200 to 900
°
C
U
-200 to 400
°
C
N
0 to 1300
°
C
R
0 to 1700
°
C
±
(0.2% of r3
°
C)
S
0 to 1700
°
C
except
±
8
°
C for 0 to 200
°
C and
±
5
°
C for 200 to 800
°
C
B
400 to 1800
°
C
±
(0.2% of r4
°
C) except
±
8
°
C for 400 to 700
°
C
W
0 to 2300
°
C
±
(0.2% of r 3
°
C)
KPvsAu7Fe
0 to 300 K
0 to 50 K :
±
8.0 K
50 to 300 K :
±
4.5 K
Maximum input voltage
(across input terminals H and L*
3
)
(At 1 kHz or less)
42 V(DC+ACpeak)(CAT I and II, 30 Vrms)
Maximum allowable common
(across input terminal H or L and earth*
4
)
mode voltage (At 1 kHz or less)
42 V(DC+ACpeak)(CAT I and II, 30 Vrms)
Input connector
Binding post
Input impedance
1 M
Ω
Common mode rejection ratio
80 dB(50/60 Hz, signal source resistance 500
Ω
or less)
120 dB(50/60 Hz, 2 Hz filter ON, signal source resistance 500
Ω
or less)
Vertical resolution
During waveform display and cursor measurements
: 0.1˚C
During automatic measurement of waveform parameters : 0.4˚C
Temperature coefficient
±
(0.02 % of FS)/
°
C
Reference junction compensation
±
1
°
C(K, E, J, T, L, U, N)
accuracy (when input terminal
±
1.5
°
C(R, S, B,W)
temperature is balanced)
±
1 K(KPvsAu7Fe)
Bandwidth limit
Select from OFF/2 Hz/8 Hz
*1 Under standard operating conditions (refer to page 15-6)
*2 Reference junction temperature compensation accuracy not included
H
L
701860
Input
terminal
earth
*4
*3