15-9
15
IM 701830-01E
Specifications
15.18 High-speed Isolation Module (701855)
Item
Specifications
Number of input channels
1
Input coupling settings
AC, DC, GND
Maximum sample rate
10 MS/s
Input format
Isolated unbalanced
Frequency characteristics*
1
DC to 1 MHz
(-3 dB point when sine wave
of amplitude
±
3 div is input)
Voltage-axis sensitivity setting
5 mV/div to 20 V/div (1-2-5 steps)
when using 700929 : 50 mV/div to 200 V/div (1-2-5 steps)
Maximum input voltage
main instrument only (across input terminals H and L
*2
)
(At 1 kHz or less)
250 V(DC+ACpeak)(CAT I and II, 177 Vrms)
combined with 700929 (across probe tips H and L
*4
)
850 V(DC+ACpeak)(CAT I and II, 600 Vrms)
Maximum allowable common
main instrument only (across input terminal L and earth
*3
)
mode voltage
42 V(DC+ACpeak)(CAT I and II, 30 Vrms)
(At 1 kHz or less)
combined with 700929 (across probe tip H or L and earth
*5
)
400 Vrms (CAT I and II)
(Specification when the EN61010-1 standard does not need to be satisfied: 600 Vrms)
DC offset range (max)
5 mV/div to 50 mV/div
:
±
0.5 V
(At 1:1 probe attenuation)
100 mV/div to 500 mV/div :
±
5 V
1 V/div to 5 V/div
:
±
50 V
10 V/div to 20 V/div
:
±
250 V*
6
Vertical (voltage) axis accuracy
DC accuracy*
1
10 mV/div to 20 V/div
:
±
(1.5% of 8 div + offset voltage accuracy)
5 mV/div
:
±
(2.0% of 8 div + offset voltage accuracy)
offset axis accuracy*
1
±
(0.04% of offset voltage range + 1% of setting value)
Input connector
BNC connector (isolating type)
Input impedance
1 M
Ω
1%, approx. 30pF
-3dB point for AC coupling used
10 Hz or less(1 Hz or less when using 700929)
Common mode rejection ratio
80 dB(50/60 Hz) or more (typical*
8
)
Residual noise*
7
Larger of
±
400
µ
V or
±
0.06 div(typical*
8
)
Withstand voltage
3700 VAC for 1 minute
(across input terminal and earth)
(60 Hz)
Isolation resistance
500 VDC, 10 M
Ω
or more
(across input terminal and earth)
A/D conversion resolution
12 bit (500 LSB/div)
Probe attenuation settings
1:1, 10:1, 100:1 and 1000:1
Bandwidth limit
Select from OFF/500 Hz/5 kHz/ 50 kHz/ 500 kHz
*1 Measured value under standard operating conditions (refer to page 15-6)