6-5
6
IM 701830-01E
Triggering
6.5
Setting the Hold off Time
≡
For a description of this function, refer to page 1-11.
≡
Function
This function prevents a trigger from being activated for a specified time, even if the trigger
conditions are met during this time. This is useful when you want to activate a trigger after a
certain repetitive period of time as illustrated in the example below.
Input signal
Repetitive period(T)
Trigger level
Trigger source
signal
t
Trigger signal during hold off period (t) (when trigger slope is set to rise)
Hold off time setting range
0 ns to 1 s (default: 0 ns)
The setting resolution is 500 ns.
Points for attention
• If holdoff time is set to 50 ms or above, the trigger mode should be set to “normal.”
• When used with the A
→
B(n) or A Delay B trigger, the holdoff time operates with respect to
condition A only.
Operating Procedure
1.
Press the
SIMPLE
key or the
ENHANCED
key.
2.
If jog shuttle action is not set to “
Hold Off
”, press the “
Hold Off
” soft key to set the action
to Hold Off.
3.
Turn the jog shuttle to set the desired hold off time.
You can change the setting a digit using the arrow keys (located below the jog shuttle).
Pressing the
RESET
key, then turning the jog shuttle sets the hold off time to 0 s.
Simple trigger mode
Enhanced trigger mode
Note
The hold off time set applies to both the simple and enhanced trigger.