9-8
IM 701830-01E
Setting the distal and proximate points (Unit)
12.
Press the “
Next
” soft key to display the “
Next 2/2
” menu.
13.
Press the “
Dist/Prox Mode
” soft key as necessary to select Unit.
14.
Press the “
Trace
” soft key to select the trace target.
15.
Press the “
Distal
” soft key to set the jog shuttle action to Distal.
16.
Turn the jog shuttle to set the distal point.
17.
Set the mesial and proximal points in the same way.
For Manual
For Unit
Setting High/Low points
12.
Press the “
Next
” soft key to display the “
Next 2/2
” menu.
13.
Press the “
High/Low Mode
” soft key to select “
Auto
” or “
Max-Min
”.
Note
• You can select up to sixteen items in total on all channels. However, Histo can be selected
independently of the sixteen items.
• The trace and item menus can be displayed at the same time. This is useful when you wish to measure
parameters on multiple channels.
• For measurement of delay (Delay Ref: CH), the mesial point next to the continuous line (vertical line) is
used as the reference point.
• If Delay Ref is set to a channel (CH1 to CH16), the measurement range extends from the Delay Ref line
to the right edge of the screen.
9.2 Automatic Measurement of Waveform Parameters