9-7
9
IM 701830-01E
Waveform Analysis
Setting the measurement range
9.
Press the “
Time Range
” soft key.
10.
Turn the jog shuttle to set the desired measurement range.
Setting the delay start point (for measurement of RDelay and FDelay)
11.
Keep pressing the “
Delay Ref
” soft key to select the desired source of the delay start point
from “
Trig
” to “
CH16
”.
“
CH1
” to “
CH16
” can be directly selected by pressing
CH1
to
CH16
.
Channels which do not have input modules installed are not displayed on the menu.
When one of the channels CH1 to CH16 is set as the Delay Ref
12.
Turn the jog shuttle to set the delay measurement start line (indicated by the continuous line
on the screen).
13.
Press the “
Delay Ref Slope
” soft key to select the slope.
Setting the distal and proximal points (Auto and Manual)
12.
Press the “
Next
” soft key to display the “
Next 2/2
” menu.
13.
Press the “
Dist/Prox Mode
” soft key to select “
Auto
” or “
Manual
”.
14.
Press the “
Distal
” soft key.
The distal point can be set using the jog shuttle.
15.
Turn the jog shuttle to set the desired distal point.
16.
In the same way, set the proximal and mesial point.
9.2 Automatic Measurement of Waveform Parameters