9-18
IM 701830-01E
9.9
Manual Scaling
≡
For a description of this function, refer to page 1-20.
≡
Function
Normally, the computed waveform is displayed using auto scaling according to the results of
the computation. Using manual scaling, you can set the upper and lower limit values of the
computed waveform display to any desired value.
The following computed waveforms cannot be manually scaled.
• Waveforms of data that are converted to binary values simultaneously on all channels.
• Waveforms of binary data that are D/A converted.
Setting Range
-1.0000E+30 to +1.0000E+30
You can read the auto scaling value by setting the “Scale Value” to ON in the DISPLAY menu.
Operating Procedure
1.
Press the
MATH
key.
2.
If “
User Define
” is selected for the “
Mode
”, display the “
Next 2/3
” menu.
If any other mode is selected display the “
Next 1/2
” menu.
3.
Press the “
Scale & Unit
” soft key to display the scaling value setting menu.
4.
Move the cursor to “
Mode
” with the jog shuttle.
5.
Press the
SELECT
key to select “
Manual
”.
→
6.
Turn the jog shuttle to move the cursor to “
Upper
” under “
Math1
”, and press the
SELECT
key to display the menu for setting the value.
7.
Set the upper limit value with the jog shuttle.
8.
Similarly, set the lower limit value at “
Lower
”.
9.
As necessary, repeat steps 4 through 8 to set “
Math2
”.
→