6-20
IM 701830-01E
Operating Procedure
Set the trigger type
1.
Press the
ENHANCED
key.
2.
Press the “
Type
” soft key to open the Trigger Type menu.
3.
Press the “
Type
” soft key as necessary to select Window.
Set the trigger condition
4.
Press the “
Set Pattern
” soft key to open the trigger condition menu.
5.
Turn the jog shuttle and move the cursor to CH“
1
” of <Window OR>.
6.
Press the
SELECT
key to display the CH“
1
” menu of <Window OR>.
7.
Press the same
SELECT
key several times to select “
,” “
,” or “-.”
Select the waveform
8.
Press the “
Select CH
” soft key to select the desired waveform from channels “
CH1
” to
“
CH16
.”
Set the window center
9.
Press the “
Center/Window Width
” soft key as necessary to set the jog shuttle action to
“
Center
”.
10.
Turn the jog shuttle to set the center level.
Set the width
11.
Press the “
Center/Window Width
” soft key as necessary to set the jog shuttle action to
“
Window Width
”.
12.
Turn the jog shuttle to set the width.
Other settings
13.
You must also set the following parameters: trigger mode, trigger hysteresis and holdoff
time. For information about how to set these parameters, refer to Sections 6.1 to 6.7.
6.14 Setting the Window Trigger (ENHANCED)