3-2
IM 701830-01E
3.2
Installing
Installation Conditions
The instrument must be installed in a place where the following conditions are met.
Flat horizontal location
Set the instrument in the proper direction and in a level and stable place. If placed in an uneven
or unstable place, printing quality decreases. Note that the instrument can be tilted on the stand.
Well-ventilated location
Vent holes are situated on the bottom side of the instrument. In addition, vent holes for the
cooling fans are also situated in the left side and the bottom. To prevent a rise in the internal
temperature, the vent holes should not be blocked and sufficient clearance should be maintained
around them.
5cm
or more
5cm
or more
When connecting the various cables or when opening or closing the built-in printer cover, take
extra space in addition to the space indicated above for the operation.
Ambient temperature and humidity
Ambient temperature: 5 to 40
°
C
Ambient humidity:
20 to 80% RH (when not using the printer)
35 to 80% RH (when using the printer)
No condensation should be allowed.
Note
• To ensure high measurement accuracy, the instrument should only be used under the following
conditions.
Ambient temperature: 23
±
5
°
C
Ambient humidity: 55
±
10% RH
• Internal condensation may occur if the instrument is moved to another place where both the ambient
temperature and humidity are higher, or if the temperature changes rapidly. In such cases allow the
instrument to acclimatize to its new environment for at least one hour before starting operation.
Never install the instrument in the following places.
In direct sunlight or near heat sources
Where an excessive amount of soot, steam, dust or corrosive gases are present.
Near magnetic field sources
Near high voltage equipment or power lines
Where the level of mechanical vibration is high
In an unstable place