17-4
IM CW240E
Formula for Each Wiring
Measurement
Item
Wiring
Single Phase 3 Wire
Equations
Symbol
Average Voltage
Uave
( U1
U2 ) / 2
Iave
( I 1
I 2 ) / 2
∑
P
P1
P2
∑
Q
Q1
Q2
Average Current
Active Power
Reactive
Power 2
*4
Apparent Power
S = U
I
∑
S
S1
S2
Power Factor
Reactive power
method not used
*6
∑
PF
Reactive power
method used
∑
PF
Phase Angle
*6
∑
PA
Q
=
S
2
P
2
∑
PF
=
∑
S
∑
P
∑
PA
= cos
–1
∑
PF
∑
P
2
∑
Q
2
∑
PF
=
∑
P
_
_
_
_
Measurement
Item
Wiring
3 Phase 3 Wire 2 Current
*7
Average Voltage
Uave
( U1
U2
U3 ) / 3
Iave
( I 1
I 2
I 3 ) / 3
*2
∑
P
P1
P3
∑
Q
Average Current
Active Power
Reactive
Power 2
*4
Apparent Power
S
U
I
∑
S
Power Factor
Reactive power
method not used
*6
∑
PF
Reactive power
method used
∑
PF
Phase Angle
*6
∑
PA
∑
PF
=
(
S
1
S
3)
∑
S
∑
P
Q
=
S
2
P
2
∑
S
2
∑
P
2
∑
P
2
∑
Q
2
∑
PF
=
∑
PA
= cos
–1
∑
PF
∑
P
2
3
Equations
Symbol
_
_
_
_
17.1 CW 240 Specifications