8-14
IM CW240E
8.4 Backup Memory
• If integration measurement has been completed when the PC card is full or
extracted, you need to take the following steps:
If the data in the backup memory is required,
copy backup data.
• Backup data can be copied on the File screen.
See Section 9.7, Backup Memory Copy.
If none of the data in the backup memory is required,
delete backup data.
• Backup data can be deleted on the File screen.
See Section 9.8, Backup Memory Delete.
• You can start new integration to delete backup data.
Starting a new integration measurement causes data in the backup memory
to be deleted.
TOP
MENU
ENTER
LIGHT
TOP
MENU
SAVE
A
RANGE
START
&STOP
ESC
ENTER
F
1
F
2
F
3
F
4
F
5
DISP COPY
RANGE
NOTE
It is recommended that even if DATA SAVE is set to PC CARD, file(s) in the
internal memory be deleted before starting an integration measurement to se-
cure the memory space for backup.
If the internal memory becomes full, data cannot be backed up in the event of a
failure to save data in a PC card.