16-5
IM CW240E
Maintenance Troubleshooting
16
Message
Maximum number in PC card
exceeded.
Content/Action
The number of saved files has exceeded 30. Delete
unnecessary files from the PC card.
Not flash ATA card.
The inserted PC card is not flash ATA. Insert a flash
ATA memory card.
PC card not supported.
An unsupported flash ATA memory card has been
inserted. Insert a supported card.
Error relating to the PC card has
occurred.
A PC card-related error other than the above has
occurred. Check the PC card.
Maximum number in setting file
exceeded.
The number of saved setting files has exceeded 30
(5 in the case of internal memory). Delete
unnecessary files.
Internal memory exceeds capacity.
The capacity of the internal memory has been
reached. Delete unnecessary files.
Maximum number in internal memory
exceeded.
The number of files saved in the internal memory has
exceeded 30. Delete unnecessary files.
No saved data in backup memory.
Backup memory copy or backup memory delete was
selected when there is no data saved in the backup
memory. This operation is not required.
Backup memory exceeds capacity.
The backup memory is full. Delete data from the
backup memory.
Correct wiring.
Confirm that connections have been made correctly,
using Scott connection confirmation.
1P3W load connection is
different from setting. It may
connect to R-S (example).
Check setup of 1P3W load
Connection.
It is possible that Scott connection confirmation is not
correct. Check the connections and the CW240
settings.
Failure in saving data within
output interval time.
It becomes easy for this problem to occur when there
are many files in the save destination or when the file
capacity is large. Check the save destination memory.
Start time has passed. Start
Integration measurement by
JUST start.
The START&STOP key was pressed after the set start
time had passed. Start integration measurement with
JUST start.
Stop time is over. End
measurement by hand.
The START&STOP key was pressed after the set stop
time had passed. Start integration measurement.
Also, stopping is done manually. Press the
START&STOP key to stop integration measurement.
The SAVE key or the DISP COPY key was pressed
during integration measurement. Perform saves and
screen copies after integration measurement has ended.
Invalid key operation during
integration measurement.
PC card exceeds the capacity.
Delete unnecessary files from the PC card.
16.2 Error Message Content and Actions