8-13
IM CW240E
Saving Measured Data
8
8.4
Backup Memory
• When the measured data-saving destination is set to a PC card, the internal
memory is used as a backup memory for the PC card.
If data cannot be written into the PC card for any reason, data will be saved in
the backup memory.
• If the backup memory contains data, the
mark is indicated on the screen.
(Note that data in the backup memory will be retained even when power is
turned OFF.)
• Data in the backup memory will be cleared each time integration
measurement is started. (If the next integration measurement is made with
data retained in the backup memory, such previous data will be deleted.)
• If the backup memory becomes full of data, an error message is displayed,
causing a save to terminate. (Integration measurement still continues.)
• Backup time varies depending on the remaining capacity of the internal
memory.
• If a PC card with free space is inserted into the CW240 during integration,
successive data will be saved in the PC card under the same file name. When
integration is complete, data that has been saved in the backup memory will
be copied to the end of the file in the PC card. (A message is displayed,
indicating that backup memory data is being copied to the PC card.)
• If no PC card with free space is inserted into the CW240 during integration, the
instrument keeps on saving data in the backup memory.