6-53
IM CW240E
Configuring Settings
6
6.4.3
Precautions for Selecting the Interval Time
Standard Interval
1 wave
100ms
200ms
500ms
1sec
2sec
5sec
10sec
15sec
30sec
1min
2min
5min
10min
15min
30min
60min
Short-time Interval
When making demand measurement, note the following:
• A demand interval of demand measurements is common to the interval time.
If the interval time is set to a value less than short-time interval (less than
1 sec.), the demand measured value cannot be displayed (the symbol "----"
appears).
When saving (outputting) data with the interval time set to 2 min or less,
note the following:
• There may be cases where data is missed at the start depending on the
number of files or space in a PC card. Set up the starting date and time with
allowances.
• As the number of data increases, there are cases where data is missed in
saved data.
• When using the CW240, especially when a short-time interval is set, note the
following:
Ensure there is no file in the PC card. (Format the PC card.)
Start integration with a PC card inserted in the CW240.
Do not remove the PC card from the CW240 during integral measurement.
Do not perform communication.
Do not operate the operation keys excessively.
If the interval time is set to 30 sec or less, note the following:
• If RS-232 CONNECT has been set to PRINTER, change the setting to PC.
(The
M
mark appears.)
• If HARD COPY has been set to PRINTER, change the setting to PC card.
• Note that automatic output of harmonic measured data or waveform data, or
automatic printout of it to a printer, is not possible.
If the interval time is set to a short-time interval, note the following:
• For interval time-basis outputs during integration measurement, only
instantaneous values of the normal measurement data are output.
• Files are generally created in a binary format. If a file is to be read by
spreadsheet software, etc., conversion to a text file is required. (For
conversion software, see the CD-ROM provided for the CW240.)
6.4 Save Data Settings 1/2