Installation
2-6
Installation
Thermo Fisher Scientific
REC 4132 Rev
E
Line power should be supplied through one of the holes provided on the
right side of the lower enclosure. The proper wiring should be capable of
providing 15A/100 VAC or 7.5A/240 VAC.
EZx with a Vortex cooler 14 AWG stranded wire should be used with
watertight metal conduit. EZx with an air conditioner 12 AWG stranded
wire should be used with watertight metal conduit. The proper wiring
should be capable of providing 20A/115VAC or 10A/230 VAC. The
branch circuit must be externally fused at no more than 20A.
The line power connection terminals for the
EZx
are located on the
wiring panel terminal block TB1. The “hot” and “neutral” conductors
must be connected to terminals L1 and L2 respectively. These terminals
have a torque specification of 0.6 N/m or 5.5 lbs/in.
A ground terminal located on the bottom left of the enclosure is
provided for safety ground connections. This terminal should be
tightened to 5.1 N/m or 45 lb/in.
Refer to the “
Back Panel Wiring Diagram
” (
Appendix F
) for detailed
information. All Line power wiring must conform to National Electric
Code (US) and/or any local regulations. The EZx incorporates universal
voltage and connects anywhere in the world.
A96733
Figure 2–3.
I/O Board
Connecting
Power to the
System
A96732
Ground
Summary of Contents for EZx
Page 1: ...EZx Contaminant Detector User Guide REC 4132 Rev E Part Number 062477...
Page 7: ......
Page 8: ......
Page 13: ...Contents Thermo Fisher Scientific Contents v REC 4132 Rev E This Page Left Blank Intentionally...
Page 14: ......
Page 112: ......
Page 214: ......
Page 228: ......
Page 231: ...This Page Left Blank Intentionally...
Page 232: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......