Maintenance and Troubleshooting
4-26 Maintenance and Troubleshooting
Thermo Fisher Scientific
REC 4132 Rev
E
A96346
Figure 4–17.
Output Diagnostics
Path:
More Menus/Test/Diode Array
The diode array test turns
X-ray
s on using the currently selected product
settings and records the
X-ray
intensity as measured by each detector
diode. The number of diodes varies with aperture size.
Large differences between adjacent diodes is indicated in red, and may
indicate a diode failure if it is in the active area. The average count in the
center region is also displayed.
Select
D
IODE
A
RRAY
and the following screen displays:
Diode Array Test