Service, Repair, and Replacement Parts
Thermo Fisher Scientific
Service Repair and Replacement Parts 5-7
REC 4132 Rev
E
Note.
If the belt is too tight to be slipped off, loosen the motor
mounting screws to release the tension on the belt.
▲
10.
Do the following to loosen the motor mount (
Figure 5–4
)
A96304
Figure 5–4.
Motor Mounting Screws
•
Use a straight tip screwdriver to loosen the screws on the motor
mount. Loosen just enough to release tension on the belt.
11.
On the opposite side of the motor, flip up the bearing retainer latch
(
Figure 5–5
).
A96305
Figure 5–5.
Bearing Retainer Latch
12.
Slide the drive roller assembly toward the pulley end and remove
from the belt.