Maintenance and Troubleshooting
4-14 Maintenance and Troubleshooting
Thermo Fisher Scientific
REC 4132 Rev
E
A96352
Figure 4–8.
Annunciation Lamp
CAUTION.
There are
X-ray
hazards inside the main cabinet.
DO
NOT
bypass any safety interlocks.
•
Check inside the machine for water ingress and the effectiveness of
the door seals.
•
Check the condition of all the fans inside the machine. Noisy or
vibrating fans should be replaced.
•
Check that the air filters are clean and free from debris. Clean or
replace as necessary. (Filters 1 & 2 are mounted on the frame).
Check roller bearings for heat, noise, or vibration. Contact
Thermo
Fisher Scientific
for replacement.
Check Roller
Bearings
Annunciation Lamp –
illuminates when X-
rays are turned on.
Failure of this lamp
will prevent X-rays
from being generated.