Maintenance and Troubleshooting
Thermo Fisher Scientific
Maintenance and Troubleshooting 4-15
REC 4132 Rev
E
A visual check after 40 minutes of work at 100% power. Open the
bottom compartment containing the tank and check for oil leaks in and
around the compartment. Contact
Thermo Fisher Scientific
immediately
if any leaks are found.
Open all compartments and check for water in and around the
compartment. Contact
Thermo Fisher Scientific
immediately if any leaks
are found.
Open the bottom cabinet and make sure fans are running. If they are not,
contact
Thermo Fisher Scientific
immediately.
Check the filters and clean or replace as necessary. If there is a drop in
air pressure, a plugged or dirty filter could be the reason. The air filters
are part of the
Filter Regulator Assembly
.
A96727
Figure 4–9.
Filter Regulator Assembly
Remove the bowls on the bottom of the
Filter Regulator
to expose the
filters.
Check for Oil Leaks
Check for Water
Ingress
Check Fans
Check Air Filters