Maintenance and Troubleshooting
Thermo Fisher Scientific
Maintenance and Troubleshooting 4-7
REC 4132 Rev
E
Belt tracking and adjustment is accomplished by tightening or loosening
the hex-head adjustment screws on the tail roller assembly while the
conveyor belt is moving.
If the belt is tracking excessively toward one side, use an 8 mm wrench
to tighten the adjustment screw on that side or loosen the adjustment
screw on the opposite side.
Maintain enough tension to prevent slippage of the conveyor belt on the
drive roller when the machine is loaded with product.
Table 4–1.
EZx
Daily Maintenance Checklist
Daily Maintenance
Completed
Clean as necessary
See Cleaning the System
Perform a Radiation Survey if required
See Appendix A
Check the condition of the shielding curtains
at the entry and exit of the hood. If they are
worn or damaged, replace them.
Check the belt tension
See page 4-4
Check the air pressure
See page 4-4
Check all photo sensors
See page 4-5
Drain fluids from the Filter Regulator Bowls
See page 4-6
After starting the conveyor
Check belt tracking and speed and adjust as
necessary
See page 4-7
Belt Tracking
Summary of Contents for EZx
Page 1: ...EZx Contaminant Detector User Guide REC 4132 Rev E Part Number 062477...
Page 7: ......
Page 8: ......
Page 13: ...Contents Thermo Fisher Scientific Contents v REC 4132 Rev E This Page Left Blank Intentionally...
Page 14: ......
Page 112: ......
Page 214: ......
Page 228: ......
Page 231: ...This Page Left Blank Intentionally...
Page 232: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......