Maintenance and Troubleshooting
Thermo Fisher Scientific
Maintenance and Troubleshooting 4-11
REC 4132 Rev
E
Path
: Main Menu/Test/X-ray Detection
On a weekly basis, check the temperature of the oil in the tank. The
temperature inside of oil tank has to be less than 67
°
C after 40 minutes
of work at 100% power.
The temperature displays at the
T
EST
M
ENU
under the
X-ray
Diagnostics
screen.
Path
: Run/select running product/Diode Calibration
Diode Calibration normalizes or evens the response of the detector
diodes, making each diode read from 0 - 100 even though all diodes
detect X-rays differently.
Note.
A Diode Calibration should be performed on a per product
basis.
▲
Diode calibration should be performed weekly. Use the following steps
to run
Diode Calibration
.
1.
Stop the running product
2.
On the front screen select
R
UN
, the following screen displays
A96723
Figure 4–5.
Diode Calibration
Check Temperature
of the Oil in the Tank
Diode Calibration