Operation
Thermo Fisher Scientific
Operation 3-9
REC 4132 Rev
E
Check the emergency stop to ensure it is working correctly. The X-rays
and the conveyor belt motor are disengaged when an emergency stop is
activated. When the E-stop switch is re-set, X-rays should not come
back on and the motor should not start until the machine is instructed to
do so using the touch screen and the motor start/stop switch.
Path
: More Menus/Test/X-ray Diagnostics
1.
As in the Critical Radiation Survey select
START
X-
RAYS
2.
After the X-rays have started, push the
EMERGENCY STOP
located
on the front of the machine and make sure the X-rays disengage.
Check that the interlocks and E-stops are working correctly. The
interlocks are placed inside the conveyor hoods and prevent X-rays from
being emitted once the hood is slightly open. The E-stop is to be used in
case of emergency to stop the x-rays and conveyor belt.
Check the
Emergency
Stop
Check
Interlocks and
E-Stop
Bottom Panel
Front Panel
Center Hood
Left and Right Hoods
E-Stop