Maintenance and Troubleshooting
Thermo Fisher Scientific
Maintenance and Troubleshooting 4-9
REC 4132 Rev
E
To test the interlocks begin at the front screen. Open the hoods and
doors individually. When a hood or door is opened a
Safety Circuit
Problem
fault message displays on the screen.
A96734
Figure 4–3.
Safety Circuit Problem
Select the error display portion of the screen and the following screen
displays:
A96735
Figure 4–4.
Safety Circuit Errors
Interlock and E-Stop
Check