EZx Menus
Thermo Fisher Scientific
EZx Menus
A-39
REC 4132
Rev E
A96368b
Figure A–38.
Diode Calibration in Manual Filter
17.
When the message at the bottom of the screen indicates the diodes
are calibrated, pass several uncontaminated products through the
hood (10 minimum).
18.
Select
STOP THRESHOLD
19.
The machine is now in verify mode. Pass both contaminated and
uncontaminated products.
20.
Select
Stop Check
21.
Select
Accept
Fine Tuning the Thresholds
If the EZx fails to detect or falsely rejects a good package during the
verify mode of the threshold learning process (see step 21 above in the
Manual Filter Setup example) select “
Change Filter Limits
”. Refer to
Table A–3
and verify you are not using a contaminant that is too small.
Remember actual sensitivity may vary depending on the characteristics
of the product, the aperture size, and the belt speed (sensitivity decreases
as the belt speed increases).
Try adjusting the threshold of the offending filter (indicated by the color
on the image) in 3% increments. If the contaminant is missed, increase
the threshold to increase sensitivity. Conversely, if the EZx is rejecting a
good package, decrease the threshold to reduce sensitivity.
Summary of Contents for EZx
Page 1: ...EZx Contaminant Detector User Guide REC 4132 Rev E Part Number 062477...
Page 7: ......
Page 8: ......
Page 13: ...Contents Thermo Fisher Scientific Contents v REC 4132 Rev E This Page Left Blank Intentionally...
Page 14: ......
Page 112: ......
Page 214: ......
Page 228: ......
Page 231: ...This Page Left Blank Intentionally...
Page 232: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......