Introduction
1-10
Introduction
Thermo Fisher Scientific
REC 4132 Rev
E
Conveyor Length
1.6 meters/5.25 feet (Compact Option: 1.0 meter/3.28 ft)
Compressed Air
Dry 80-100 PSI (5.5-6.9 bar), 30 - 40 SCFM (1135 SLPM)
See below for required air line sizing.
Air Line I.D.
Distance from Supply
9.5mm (3/8”)
0 - 10 ft (0 - 3 m)
12.7mm (1/2”)
10 - 50 ft (3 - 15 m)
15.9mm (5/8”)
50 - 100 ft (15 - 30 m)
Do not use restrictive fittings such as quick connects. They can “starve”
the Vortex cooler by causing an excessive line pressure drop.
Environmental
Storage
Ambient Air Temperature: -25 to 70° C (-13 to 158° F)
(For <24 hours)
Rate of Change of Temperature: 5° C/min (9°F/min)
Relative Humidity: 5 to 95% non-condensing
Absolute Humidity: 0.1 to 35 g/m
3
Operating Environment
Ambient Air Temperature: +15 to +40°C (60 to +104°F)
X-ray
Tank Body Temperature: +15 to +61°C (59 to +136°F)
Maximum rate of Temperature Change: 0.5 °C/min (41°F/min)
Relative Humidity: 20 to 90% non-condensing
Absolute Humidity 0.1 to 35%
Condensation: IP65 Washdown
Pollution Degree
2 (Normally only non-conductive pollution occurs. Occasionally a
temporary conductivity caused by condensation may occur.)
Altitude
Up to 2000 m (6560 ft)
Summary of Contents for EZx
Page 1: ...EZx Contaminant Detector User Guide REC 4132 Rev E Part Number 062477...
Page 7: ......
Page 8: ......
Page 13: ...Contents Thermo Fisher Scientific Contents v REC 4132 Rev E This Page Left Blank Intentionally...
Page 14: ......
Page 112: ......
Page 214: ......
Page 228: ......
Page 231: ...This Page Left Blank Intentionally...
Page 232: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......