Maintenance and Troubleshooting
Thermo Fisher Scientific
Maintenance and Troubleshooting 4-21
REC 4132 Rev
E
The
EZx
software is equipped with a fault diagnostic to help find
specific faults. If the software detects a fault in the
X-ray
safety system,
the
X-ray
s are automatically shut down.
CAUTION.
Only personnel trained by
Thermo Fisher Scientific
or
an agent acting on behalf of
Thermo Fisher Scientific
should
attempt to correct faults on the
EZx
or damage to the system
could result.
▲
Although it is advised that only trained personnel should attempt to fix
faults on the
EZx
there are several quick checks the operator should
make to avoid delays.
Check the screen display
1.
If the display screen is blank check power to the system as described
below.
2.
The
EZx
takes several minutes to “boot-up”. If after several minutes
the system is not at its initial screen, turn the Main switch to the
OFF
position, wait for a few seconds and then turn it to the
ON
position.
Use the following steps to check the power to the system.
1.
Make sure power is
ON
at the mains
CAUTION.
Only appropriately trained personnel should open the
main cabinet.
▲
2.
Check all circuit fuses in the machine
•
Fuse F2 – Electronics line power
•
Fuse F3 – Conveyor Motor line power
•
Fuse F1 – Ancillary devices line power
•
Fuse F4 – 24 VDC bus
Troubleshooting
System is not
Operating
Check the Power to
the System
Summary of Contents for EZx
Page 1: ...EZx Contaminant Detector User Guide REC 4132 Rev E Part Number 062477...
Page 7: ......
Page 8: ......
Page 13: ...Contents Thermo Fisher Scientific Contents v REC 4132 Rev E This Page Left Blank Intentionally...
Page 14: ......
Page 112: ......
Page 214: ......
Page 228: ......
Page 231: ...This Page Left Blank Intentionally...
Page 232: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......