Operation
Thermo Fisher Scientific
Operation 3-19
REC 4132 Rev
E
4.
When the message at the bottom of the screen indicates the diodes
are calibrated, pass several uncontaminated products through the
hood (10 minimum).
5.
Select
S
TOP
T
HRESHOLD
6.
The machine is now verifying. Pass both contaminated and
uncontaminated products. Observe whether the current filter settings
can detect contaminated product without false tripping on good
product. If contaminated product is detected, go to the Step 6.
If contaminated product is not detected or the machine is false
tripping the filters may require manual adjustment. Refer to Manual
Filter on page A-33.
7.
Select
S
TOP
C
HECK
8.
Select
A
CCEPT
Hint.
The more products passed at this time the more reliable the
threshold results.
▲