Maintenance and Troubleshooting
Thermo Fisher Scientific
Maintenance and Troubleshooting 4-19
REC 4132 Rev
E
Type
Cause
Recovery
Motor Speed Fault
No speed pulses or incorrect
speed pulses
•
Turn on motor
•
Check
motor
controller
•
Check
belt
Air Pressure Low
Air pressure going to the
reject device is low.
•
Fix air pressure problem
Invalid Diode Count
EZx detects a different
number of diodes than
entered in global setup
•
Check Global setup parameter, by
selecting aperture size
•
Perform diode array test to see if any
diodes are bad
Cold Start
Database has been deleted
Clear alarm, this alarm is to make sure the
operator acknowledges that the system has
been cold started
Type
Cause
Recovery
Bad Data base
Unable to read or write to
database
Cold start the machine
Start-up Failure
Variable required for
interface thread invalid;
Unable to open control
thread.
•
Check USB connection between Host &
IPU board
•
Cold Start the machine
Power Supply
X-ray
Fault
•
Open
control
loop
•
Invalid run command
•
Invalid KV or mA set
point
•
mA
unstable
•
KV balance mismatch
•
Arc-over
Some of these conditions may be recoverable
others may not. Contact
Thermo Fisher
Scientific Service
if the fault persists.
Power Supply Tank
Fault
•
Filament over current
•
Filament open circuit
•
Oil over temp
•
Temp
sensor
open
•
Stir fan under speed
•
X-ray
power limit
Some of these conditions may be recoverable
others may not. Contact
Thermo Fisher
Scientific Service
if the fault persists.
Summary of Contents for EZx
Page 1: ...EZx Contaminant Detector User Guide REC 4132 Rev E Part Number 062477...
Page 7: ......
Page 8: ......
Page 13: ...Contents Thermo Fisher Scientific Contents v REC 4132 Rev E This Page Left Blank Intentionally...
Page 14: ......
Page 112: ......
Page 214: ......
Page 228: ......
Page 231: ...This Page Left Blank Intentionally...
Page 232: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......