EZx Menus
Thermo Fisher Scientific
EZx Menus
A-35
REC 4132
Rev E
Use an uncontaminated package in all of the following steps unless the
instructions specify otherwise.
Setup all other parameters of the new product.
Use the following steps to perform a
MANUAL FILTER
setup:
1.
To perform a manual filter calibration, select
Manual Filter
from the
S
ETUP
N
EW
P
RODUCT
screen.
2.
Select “
Side Edge Mask
” for the first filter. Set the thickness to 1
mm.
3.
Select “
Simple
” for the second filter.
4.
Select “
Enhanced Gradient
” for the third filter.
5.
Select “
Unassigned
” for all other filters.
Manual Filter Setup
Figure A–32.
Filter Setup
6.
Start the belt
Step I.
Set the filter
sequence to Side
Edge Mask, Simple
Threshold and
Enhanced Gradient
in that order