Installation
Thermo Fisher Scientific
Installation 2-5
REC 4132 Rev
E
The
EZx
machine is equipped with adjustable feet for adjusting the line
height to that of the mating conveyors. Use the following steps to adjust
the height:
1.
Loosen the jam nut on the bottom of any leg requiring height
adjustment using a 24 mm wrench.
2.
Use a 13 mm wrench on the flats at the bottom of the threaded
portion of each leveling foot and adjust up or down as necessary.
3.
Retighten all jam nuts.
A96350
Figure 2–1.
Adjustable Leg
4.
Ensure the system is level and vibration free.
Connect an oil free/dry compressed air line to the inlet port (1/4 NPT) of
the filter regulator on the EZx. The air supply should be capable of
delivering 30 - 40 SCFM (Standard Cubic Feet per Minute) at 80 – 100
PSI (Per Square Inch) minimum.
3
A96746
Figure 2–2.
Air Inlet
3
The reference to air does not apply if the system has an air conditioning unit.
Adjusting the
Height
Connecting Air to
the System
Summary of Contents for EZx
Page 1: ...EZx Contaminant Detector User Guide REC 4132 Rev E Part Number 062477...
Page 7: ......
Page 8: ......
Page 13: ...Contents Thermo Fisher Scientific Contents v REC 4132 Rev E This Page Left Blank Intentionally...
Page 14: ......
Page 112: ......
Page 214: ......
Page 228: ......
Page 231: ...This Page Left Blank Intentionally...
Page 232: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......