C-Link Protocol Commands
Datalogging
Thermo Fisher Scientific
Model 146
i
Instruction Manual
B-21
malloc lrec
malloc srec
These commands report the currently set memory allocation for long and
short records in percent of total memory.
Send:
malloc
lrec
Receive:
malloc lrec 10%
set malloc lrec
value
set malloc srec
value
value
= 0 to 100
These commands set the percent of memory space allocated for long and
short records to
value
, where
value
is a floating-point number representing
percent. The example below sets the memory allocation for long records to
10.
Note
Issuing these commands will clear all the logging data memory. All
the existing records should be retrieved using appropriate commands, if
required.
▲
Send:
set malloc lrec 10
Receive:
set malloc lrec 10 ok
set copy sp to lrec
set copy sp to srec
set copy sp to stream
These commands copy the current selections in scratch pad (sp) into the
long record, short record, or streaming data list. The example below copies
the current list in scratch pad into the long records list.
Send:
set copy sp to lrec
Receive:
set copy sp to lrec ok
set copy lrec to sp
set copy srec to sp
set copy stream to sp
These commands copy the current contents of the long record, short
record, or streaming data list into the scratch pad (sp). These commands
are useful in easy modification of current long record, short record, or
streaming data lists. The example below copies the current list of long
records into the scratch pad.
Send:
set copy lrec to sp
Receive:
set copy lrec to sp ok
Summary of Contents for 146i
Page 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Page 4: ......
Page 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 148: ......
Page 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Page 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Page 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...