Optional Equipment
Permeation Oven
9-4
Model 146
i
Instruction Manual
Thermo Fisher Scientific
3. Remove the thermistor from POJ1 on the measurement interface board.
4. Connect a resistance of approximately 3.8 K
Ω
across pins 1 and 2 of
POJ1.
5. Go to the Factory Cal Gas Therm menu from the Perm Oven Settings
menu in the Service menu. Select
Low Point
, enter the exact value of
the resistor and press
to calibrate the low resistance point.
6. Press
to return to the Factory Cal Gas Therm menu.
7. Connect a resistance of approximately 5 K
Ω
across pins 1 and 2 of
POJ1.
8. Go to the High Point screen, enter the exact value of the resistor and
press
to calibrate the high resistance point.
9. Press
to return to the Factory Cal Gas Therm menu.
10. Remove resistor from POJ1 and re-attach the gas thermistor.
11. Connect a resistance of approximately 3.8 K
Ω
across pins 3 and 4 of
POJ3.
12. Go to the Factory Cal Oven Therm menu from the Perm Oven Settings
menu in the Service menu. Select
Low Point
, enter the exact value of
the resistor and press
to calibrate the low resistance point.
13. Press
to return to the Factory Cal Oven Therm menu.
14. Connect a resistance of approximately 5 K
Ω
across pins 3 and 4 of
POJ3.
Summary of Contents for 146i
Page 1: ...Model 146i Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 30Jan2008 ...
Page 4: ......
Page 16: ...Contents xii Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 148: ......
Page 188: ...Troubleshooting Service Locations 6 26 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 258: ...System Description I O Components 8 16 Model 146i Instruction Manual Thermo Fisher Scientific ...
Page 273: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 15 Figure 9 5 Bench Mounting ...
Page 274: ...9 16 Model 146i Instruction Manual Thermo Fisher Scientific Figure 9 6 EIA Rack Mounting ...
Page 275: ...Thermo Fisher Scientific Model 146i Instruction Manual 9 17 Figure 9 7 Retrofit Rack Mounting ...